Chevrolet Trax: Evaporative emission control system diagnosis
Special Tools
EN 41413-VLV EVAP Service Port Vent Fitting
Operating Instructions for the Evaporative Emission System Tester (EEST)
Vehicle Setup
WARNING: Refer to Fuel and Evaporative Emission Pipe Warning .
CAUTION: Refer to Clean, Dry, Low Pressure Gas Source Caution
NOTE: When servicing the EVAP system, test pressure must not exceed 13
in H20
(0.5 PSI) as pressure in excess of this could cause system components to
perform inaccurately.
NOTE: A large difference between the vehicle temperature and shop
temperature
will seriously affect the accuracy of the tests. Always allow at least 15 min
for the vehicle temperature to adjust to the shop temperature. Refer to
Temperature Variation Instructions for more information about vehicle
versus shop temperatures.
- Engine OFF, open the hood. Position a large fan to blow air under the
vehicle onto the fuel tank area.
- Connect the red battery clip from the tester to the positive
battery terminal.
NOTE: The vehicle battery must be fully charged for optimum tester
performance.
- Connect the black battery clip from the tester to chassis ground.
Flow Meter Test - Leak Detection
CAUTION: Refer to Fuel and Evaporative Emission Hose/Pipe Connection
Cleaning
Caution
- Open the Nitrogen tank valve and turn the NITROGEN/SMOKE valve on the
front control panel to
NITROGEN.
- Connect the hose to the correct test orifice on the bottom front of the
tester. For orifice size, refer to the
vehicle specific information found in service procedures for DTCs that
relate to evaporative emission
(EVAP) system leaks.
- Press and release the remote switch to activate the tester.
NOTE: The tester will time out after about 15 min. You may press
the remote
switch to reactivate the tester as desired.
- Position the sliding red flag on the flow meter to align with the
floating indicator. When the red flag is
set, press and release the remote switch to deactivate the tester.
- Remove the hose from the test orifice and install the hose onto the
vehicle. For proper connection
location, and the special tool numbers for any adapters that may be required,
refer to the service
procedures for DTCs that relate to EVAP system leaks.
- Ignition ON, seal the EVAP system per instructions in the service
procedures for DTCs that relate to
EVAP system leaks. Most systems can be sealed using a scan tool output
control for the EVAP Vent
Solenoid Valve, other systems require that the vent system be plugged.
NOTE:
- Larger volume fuel tanks, and/or those with lower fuel levels, may
require several minutes to fill with nitrogen.
- Static buildup may cause the float indicator to stick. It may be
necessary to tap on the flow meter to free up the float.
- If fuel level is 90 % or greater it will take longer to fill the
system with
smoke because the fuel fill tube check valve will be closed and force any
smoke to pass through the ORVR pipe and orifice.
- Press and release the remote switch to activate the nitrogen flow and
fill the system.
- Compare the stable floating indicator position to the red flag.
- Below the red flag, the result is acceptable, Pass. Test complete
- Above the red flag, the result is unacceptable, Fail. Go to the Smoke
Procedure
- Press and release the remote switch to deactivate the tester.
Smoke Procedure - Locate the Leak
NOTE: It is not recommended to use the tester in an outside
repair bay area
because wind and sunlight may affect temperature and your ability to see
the smoke.
- Turn OFF any fans that may cause air movement around or near the vehicle.
NOTE: Completely unwind the nitrogen/smoke hose from the bracket to
optimize
the tester's performance.
- Connect the nitrogen/smoke hose to the vehicle as directed in the
service procedures. Some vehicles
require that the nitrogen/smoke hose be connected at the front of the EVAP
system at the EVAP service
port. An adapter may be necessary. Other vehicles require the connection be
made at the rear of the
system using an adapter at the fuel fill cap. Consult the service procedures
for DTCs that relate to
evaporative emission (EVAP) system leaks for vehicle specific instructions
regarding connection location
and adapters.
- Open the Nitrogen tank valve and turn the nitrogen/smoke valve on the
control panel to smoke.
NOTE: The remote switch operates on a push ON, push OFF
fashion.
- Press and release the remote switch to activate the tester and inject
smoke into the EVAP system.
- Verify smoke has filled the EVAP system by opening the system opposite
the end where smoke is
injected. When injecting smoke at the service port, remove the fuel fill cap
until smoke is observed. Then
close the system and continue testing. If using a special tool fuel fill cap
adapter at the filler neck, vent the
system at the service port, with special tool EN 41413-VLV EVAP Service Port
Vent Fitting, until smoke
is observed then remove the vent fitting tool and close the system to
continue with the test.
NOTE: Inject smoke in less than 2 min cycles for optimum
tester performance.
- Press and release the remote switch to deactivate the tester.
NOTE: For optimum visual smoke performance, deactivate the smoke
flow and
allow the system pressure to drop. Allowing the smoke to exit small holes
at a low flow rate greatly enhances visibility.
- Introduce smoke into the system for an additional 60 s. Continue
introducing smoke at 15 s intervals, as
necessary.
- Using a high-intensity white light, inspect the entire EVAP system path,
and look for the source of the
leak indicated by the exiting smoke. Introduce smoke at 15 s intervals, as
needed, until leak source is
identified.
Temperature Variation Instructions
The Concern
Ideal circumstances for conducting the EVAP flow meter test require equal
temperatures between the Nitrogen
gas and the vehicle EVAP system. Significant differences in temperature between
them can result in a flow or
pressure change during testing, causing misleading results. Typically, the EEST
is stored indoors,
approximately 21ºC (70ºF). Vehicles brought in for diagnosis may have an EVAP
system at significantly
different temperatures.
For Example
NOTE: With no temperature difference between the Nitrogen gas and
EVAP system, the
resulting vehicle EVAP system pressure will remain stable at 3.2 kPa, 13 in. H2O
(0.5 PSI) once pressurized, providing no leaks are present.
When the EVAP flow meter tests are performed with significant differences in
temperature between the
Nitrogen gas and the vehicle EVAP system, the following results can occur:
- An increase in flow during the flow meter test can be caused by a
vehicle's warm EVAP system cooling
down.
- A decrease in flow during the flow meter test can be caused by a
vehicle's cool EVAP system warming
up.
The Solution
When working on a vehicle with significant temperature differences between
the vehicle EVAP system and the
Nitrogen gas, allow the vehicle EVAP system temperature to stabilize as close as
possible to the temperature of
the Nitrogen gas before conducting the Flow Meter Test.
READ NEXT:
Diagnostic Instructions
Perform the Diagnostic System Check - Vehicle prior to using this
diagnostic procedure.
Review Strategy Based Diagnosis for an overview of the diagnostic
approach.
Diag
Diagnostic Instructions
Perform the Diagnostic System Check - Vehicle prior to using this
diagnostic procedure.
Review Diagnostic System Check - Vehicle for an overview of the
diagnostic approa
Diagnostic Instructions
Perform the Diagnostic System Check - Vehicle prior to using this
diagnostic procedure.
Review Strategy Based Diagnosis for an overview of the diagnostic
approach.
Diag
SEE MORE:
Introduction
The inflatable restraints passenger presence system is a calibrated system
that requires re-zeroing or a zero
check anytime the seat cushion trim attachments have been removed or the
passenger presence system has been
replaced. The procedures below are designed to assist in the re-zer
Diagnostic Instructions
Perform the Diagnostic System Check - Vehicle prior to using this
diagnostic procedure.
Review Strategy Based Diagnosis for an overview of the diagnostic
approach.
Diagnostic Procedure Instructions provides an overview of each
diagnostic category.
DTC Descriptor
DT